A novel multifocal electroretinography stimulus for detecting hydroxychloroquine retinal toxicity
My Session Status
Authors: Adrian C. Tsang, Gianni Virgili, Ange-Lynca Kantungane, Chloe Gottlieb, Stuart Coupland
Author Disclosure Block: A.C. Tsang: None. G.
Virgili: None. A. Kantungane: None. C.
Gottlieb: None. S. Coupland: None.
Abstract Body:
Purpose: To evaluate a novel 5-ring multifocal
electroretinogram (mfERG) stimulus as a screening test for detecting
hydroxychloroquine (HCQ) retinopathy.
Study Design: A stratified case-control study conducted in
accordance with the Standards for Reporting of Diagnostic Accuracy Studies
(STARD).
Methods: Consecutive patients referred to the University of Ottawa
Eye Institute for HCQ retinopathy screening from July 2018 to September 2018
underwent testing using a novel 5-ring mfERG stimulus and the standard
61-hexagon mfERG stimulus. Patients with amblyopia, high myopia or hyperopia,
coexisting retinal disease, and prior retinal surgery were excluded. mfERG
parameters were compared between protocols and against cumulative dose (CD),
dose by real body weight (RBW), and duration of HCQ therapy. Ring amplitudes
(R1-R5) and ring ratios (R1-R4/R5) were collected for each stimulus protocol.
mfERG data was collected from an additional group of controls with no previous
exposure to HCQ. Stata 15.1 software was used to perform the regression
analyses, and to compute Spearman correlation coefficients between variables.
Results: 24 eyes (10 patients, 2 controls) were included in the
final analysis. The novel R2/R5 and, R2 and R3 amplitudes, were moderately
correlated with the respective R2/R5 (r= 0.422, p=0.040), and R2 (r= 0.434,
p=0.034) and R3 (r= 0.429, p=0.036) amplitudes of the 61-hexagon stimulus. The
novel R2 P1 amplitude was highly correlated with CD (r=-0.687, p<0.001),
treatment duration (r=-0.687, p<0.001), and dose by RBW (r=-0.763, p<
0.001). Similarly, novel R2/R5 was highly correlated with CD (r= -0.743, p<
0.001), treatment duration (r= -0.743, p< 0.001), and dose by RBW (r=
-0.567, p=0.004). The correlations between the 61-hexagon stimulus R2/R5 and
HCQ exposure risk factors also reached significance, but were not as robust.
Linear regression analysis showed that CD may account for almost half of the
variance in the novel R2/R5. (r2= 0.486, R2/R5 = 9.68063 + CD*-0.00286)
Conclusions: The 2016 AAO guidelines focused primarily on the use
of subjective functional and objective structural testing, and relegated mfERG
based on a lack of accessibility. This pilot study describes a new 5-ring mfERG
protocol specific for HCQ retinopathy screening and suggests equivalence to the
61-hexagon protocol at characterizing parafoveal function. This novel 5-ring
stimulus has a significantly shorter acquisition period and the ring design may
be more sensitive to HCQ induced electrophysiologic change in the parafoveal
region. R2 P1 amplitude and R2/R5 ring ratios acquired using this novel mfERG
stimulus are strongly correlated to treatment duration and HCQ dosing.
Additional prospective cohort studies are needed to validate this novel
stimulus which may decrease the resource burden associated with objective
functional testing.